FDA Adverse Event Injury Summary report: N

LOGIC TIBIA PS MOD INSRT SZ 4 9MM

MDR report key: 19010409 · Received April 1, 2024

Report

Report Number
1038671-2024-00700
Event Type
Injury
Date Received
April 1, 2024
Date of Event
May 2, 2023
Report Date
October 22, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862001887
PMA / PMN Number
K033883
Removal / Correction Number
Z-0021-2022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10. D10. CONCOMITANTS: 1665607 02-010-01-0340 - LOGIC FEMORAL PS CEM RIGHT SZ 4 1890765 02-012-41-4030 - LOGIC TIBIA TRAPTRAY CEM SZ 4F/3T 1910639 200-02-35 - THREE PEG PATELLA 35MM THESE DEVICES ARE USED FOR TREATMENTS, NOT DIAGNOSIS. THERE IS NO OTHER INFORMATION AVAILABLE. PENDING INVESTIGATION.

Additional Manufacturer Narrative · 0

H6: ADDED THE FOLLOWING: TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. THE REASON THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR AND LOOSENING OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, THE DEVICES WERE IMPLANTED FOR OVER TEN YEARS PRIOR TO THE REPORTED FAILURE(S). HOWEVER, THE REPORTED PROSTHESIS WEAR AND LOOSENING COULD NOT BE CONFIRMED AND POTENTIAL CONTRIBUTIONS OF USER OR PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND NO IMAGES, RADIOGRAPHS, OR RELEVANT CLINICAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A RIGHT TOTAL KNEE ARTHROPLASTY ON (B)(6) 2011, AND THEN EXPERIENCED REVISION SURGICAL PROCEDURE ON (B)(6) 2023 APPROXIMATELY 12 YEARS AND 4 MONTHS AFTER INITIAL IMPLANT. NO IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1518183 LOGIC TIBIA PS MOD INSRT SZ 4 9MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862001887

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention SEE H10.