FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 1900887 · Received November 9, 2010

Report

Report Number
1824206-2010-10944
Event Type
Malfunction
Date Received
November 9, 2010
Date of Event
August 5, 2010
Report Date
October 15, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE SCALE BOARD AND THE BED SCALE FUNCTION PROPERLY.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE SCALE DISPLAY WAS FLASHING ERROR 2, CALIBRATION ERROR. THE TECH DISCOVERED THAT THE SCALE BOARD HAD SIGNS OF FLUID INGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1