COLLEAGUE CXE VOLUMETRICINFUSION PUMP
Report
- Report Number
- 6000001-2010-04916
- Event Type
- Malfunction
- Date Received
- November 16, 2010
- Date of Event
- October 1, 2010
- Report Date
- October 6, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION DISCOVERED AND CONFIRMED THE REPORTED CONDITION OF FAILURE CODE 500:320:654:0000. THE ROOT CAUSE COULD WAS DETERMINED TO BE THE START KEY BEING HELD FOR GREATER THAN 50 SECONDS BY THE USER. NO REPAIRS WERE NECESSARY TO REPAIR THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 500:320:654:0000. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THAT THIS CONDITION INTERRUPTED DELIVERY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CXE VOLUMETRICINFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |