FDA Adverse Event Malfunction Summary report: N

SPECIALITE

MDR report key: 1900652 · Received November 9, 2010

Report

Report Number
1824206-2010-10924
Event Type
Malfunction
Date Received
November 9, 2010
Date of Event
October 11, 2010
Report Date
October 13, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FSY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNTS MAINTENANCE REPLACED THE P645 SPECIALTIES' LAMP HOLDER AND HAS PLACED THE LIGHT BACK INTO SERVICE. HILL-ROM HAS SEEN SIMILAR OVERHEATING CONDITIONS IN 2-PIN, STRAIGHT LAMP FLUORESCENT LIGHTING FIXTURES. THE NATURE OF THIS TYPE OF EVENT SUGGESTS THIS CONDITION COULD OCCUR IN ANY OF THIS TYPE OF FLUORESCENT LIGHTING PRODUCTS, NOT JUST THOSE PROVIDED BY HILL-ROM. POTENTIAL CAUSES ARE: IF A 2-PIN FLUORESCENT LAMP (BULB) IS NOT SECURELY SEATED IN THE LAMP HOLDER, ELECTRICAL ARCING CAN OCCUR BETWEEN THE PINS OF THE FLUORESCENT LAMP AND THE LAMP HOLDER. IF THE ELECTRICAL ARCING IN SUSTAINED FOR AN EXTENDED DURATION, THE LAMP HOLDER MAY OVERHEAT AND POTENTIALLY IGNITE. THE RISK OF SUSTAINED ELECTRICAL ARCING IS GREATER FOR FLUORESCENT LIGHT FIXTURES THAT USE: PARALLEL BALLASTS INSTEAD OF SERIAL BALLASTS. ELECTRICAL BALLASTS INSTEAD OF MAGNETIC BALLASTS. T12 LAMPS INSTEAD OF T8 LAMPS. OUR INVESTIGATION ALSO SUPPORTS THAT A PROPERLY SEATED LAMP IN THE FIXTURE WILL NOT HAVE THIS POTENTIAL HAZARD REGARDLESS OF WHICH LAMP OR BALLAST IS USED. HILL-ROM HAS INFORMED ITS CUSTOMER BASE TO ENSURE PROPER INSTALLATION OF BULBS INTO FIXTURES.

Description of Event or Problem · 1

THE CUSTOMER ALLEGED, THEY HAD AN INCIDENT WITH WIRING BURNING ON THE INSIDE OF THE P645 LIGHT. THE INCIDENT IN RM 3616 HAPPENED OVER THE WEEKEND. NOTHING IS KNOWN AS TO IF THE ROOM WAS OCCUPIED AT THE TIME OF THE EVENT. NO INJURY WAS REPORTED. THE TECH WAS ABLE TO INSPECT THE LIGHT AND HOLDERS. HE FOUND THE HOLDERS WERE SMOKED BROWN IN COLOR AND THERE WAS SOME MELTING AROUND THE CENTER OF THE LAMP HOLDERS AND THE BACKS OF THE HOLDERS WERE OPEN. THE ACCOUNT HAS REPAIRED THE LIGHT AND PLACED IT BACK INTO SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECIALITE EXAM LIGHT FSY HILL-ROM, INC. 645

Patients

Seq Age Sex Outcome Treatment
1