FDA Adverse Event Malfunction Summary report: N

NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER

MDR report key: 19005809 · Received March 29, 2024

Report

Report Number
2029046-2024-01063
Event Type
Malfunction
Date Received
March 29, 2024
Date of Event
February 29, 2024
Report Date
April 25, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
OAD
UDI-DI
10846835000559
PMA / PMN Number
P030031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER, INC. OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 3-APR-2024, THE BWI PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE PRODUCT INVESTIGATION WAS SUBSEQUENTLY COMPLETED. IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER AND THE DEVICE (INCLUDING PORT, LUER HUB) WAS NOT IRRIGATING MID-PROCEDURE. THERE WERE NO ERRORS OR INCORRECT SETTINGS NOTED ON THE PUMP AND GENERATOR. A SECOND CATHETER WAS USED TO COMPLETE THE OPERATION. THERE WAS NO ADVERSE EVENT REPORTED ON PATIENT DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. VISUAL INSPECTION AND PATENCY TEST OF THE RETURNED DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED NO DAMAGE OR ANOMALIES ON THE DEVICE. A PATENCY TEST WAS PERFORMED, AND THE DEVICE WAS FLUSHING CORRECTLY, NO OBSTRUCTED HOLES WERE OBSERVED. NO IRRIGATION ISSUES WERE OBSERVED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 31003886M NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. THE IRRIGATION ISSUE REPORTED BY THE CUSTOMER COULD NOT BE REPLICATED DURING THE PRODUCT INVESTIGATION; OTHER ISSUES OR CIRCUMSTANCES MAY HAVE OCCURRED DURING THE USAGE OF THE DEVICE THAT COMPROMISED ITS PERFORMANCE. THE INSTRUCTIONS FOR USE (IFU) CONTAIN THE FOLLOWING RECOMMENDATIONS: FLUSH THE CATHETER AND THE TUBING TO ENSURE PURGING OF TRAPPED AIR BUBBLES AND TO VERIFY THAT THE IRRIGATION HOLES ARE PATENT. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER AND THE DEVICE (INCLUDING PORT, LUER HUB) WAS NOT IRRIGATING MID-PROCEDURE. THERE WERE NO ERRORS OR INCORRECT SETTINGS NOTED ON THE PUMP AND GENERATOR. A SECOND CATHETER WAS USED TO COMPLETE THE OPERATION. THERE WAS NO ADVERSE EVENT REPORTED ON PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2592970 NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER OAD BIOSENSE WEBSTER INC 31003886M 10846835000559

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown