HEARTMATE 3 LVAS IMPLANT KIT
Report
- Report Number
- 2916596-2024-01786
- Event Type
- Death
- Date Received
- March 29, 2024
- Date of Event
- April 1, 2023
- Report Date
- August 27, 2025
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- UDI-DI
- 00813024013297
- PMA / PMN Number
- P160054
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION CONCLUSION: A DIRECT CORRELATION BETWEEN THE HEARTMATE (HM) 3 LEFT VENTRICULAR ASSIST SYSTEM (LVAS) AND THE REPORTED PATIENT DEATHS COULD NOT BE CONCLUSIVELY DETERMINED THROUGH THIS INVESTIGATION. THE HEARTMATE 3 DEVICE SERIAL NUMBERS AND OTHER SPECIFIC CASE/PATIENT INFORMATION ARE NOT AVAILABLE AND WERE NOT REQUESTED. NO PRODUCT WAS EVALUATED UNDER THIS COMPLAINT. THE HEARTMATE 3 LVAS IFU, REV. C, IS CURRENTLY AVAILABLE. SECTION 1 OF THE IFU, ¿INTRODUCTION¿, LISTS POTENTIAL ADVERSE EVENTS, INCLUDING DEATH, THAT MAY BE ASSOCIATED WITH THE USE OF THE HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
SECTIONS A AND D: SPECIFIC PATIENT INFORMATION AND DEVICE SERIAL NUMBER ARE DOCUMENTED AS UNKNOWN. DETAILS ARE LISTED IN THE ATTACHED ARTICLE. DEVICE WAS IMPLANTED AT TIME OF EVENT. SECTION B2: SPECIFIC DEATH DATE IS DOCUMENTED AS UNKNOWN. DETAILS ARE LISTED IN THE ATTACHED ARTICLE. SECTION B3: DATE OF EVENT HAS BEEN ENTERED AS THE DATE OF PUBLICATION 01APR2023 SINCE THE DATE OF DATA COLLECTION WAS NOT PROVIDED. AUTHOR INFORMATION: MOLINA, E. J., ET AL. (2024). LEFT VENTRICULAR DIMENSIONS AND CLINICAL OUTCOMES WITH A FULLY MAGNETICALLY LEVITATED LEFT VENTRICULAR ASSIST DEVICE. JACC. HEART FAILURE, S2213-1779(24)00779-0. ADVANCE ONLINE PUBLICATION. HTTPS://DOI.ORG/10.1016/J.JCHF.2024.09.019. (B)(6) GEORGIA, USA. (B)(6) FLORIDA, USA. (B)(6) USA. (B)(6) COLORADO, USA. (B)(6) NEW YORK, USA. (B)(6) ILLINOIS, USA. (B)(6) USA. NO FURTHER INFORMATION WAS PROVIDED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER¿S INVESTIGATION IS COMPLETED.
MANUFACTURER'S INVESTIGATION CONCLUSION: A DIRECT CORRELATION BETWEEN THE HEARTMATE (HM) 3 LEFT VENTRICULAR ASSIST SYSTEM (LVAS) AND THE REPORTED EVENTS COULD NOT BE CONCLUSIVELY DETERMINED THROUGH THIS INVESTIGATION. THE SERIAL NUMBERS OR OTHER IDENTIFYING INFORMATION OF THE PRODUCTS WERE NOT REPORTED AND WERE NOT ABLE TO BE DETERMINED DURING THE INVESTIGATION. THE CURRENT REVISIONS OF THE INSTRUCTIONS FOR USE (IFU) AND PATIENT HANDBOOK CAN BE FOUND ON THE EIFU PAGE OF THE ABBOTT WEBSITE. THE HEARTMATE 3 LVAS IFU REV. D AND THE HEARTMATE 3 LVAS PATIENT HANDBOOK, REV. D ARE CURRENTLY AVAILABLE. SECTION 1 OF THE IFU, ¿INTRODUCTION¿, LISTS POTENTIAL ADVERSE EVENTS, INCLUDING DEATH, THAT MAY BE ASSOCIATED WITH THE USE OF THE HEARTMATE 3 LEFT VENTRICULAR ASSIST SYSTEM. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
RELATED MFR # 2916596-2024-01785. IT WAS REPORTED THROUGH THE RESEARCH ARTICLE ¿THE IMPACT OF SMALL LEFT VENTRICULAR DIMENSION ON OUTCOMES AFTER HEARTMATE 3 LVAD IMPLANTATION¿ THAT HEARTMATE 3 (HM3) MAY BE ASSOCIATED WITH DEATH. THIS RETROSPECTIVE STUDY COMPARED 2-YEAR SURVIVAL AND ADVERSE EVENTS OF THE 1921 PATIENTS IMPLANTED WITH HM3 WITHIN THE MOMENTUM 3 TRIAL PORTFOLIO. THE PATIENTS WERE COMPARED BASED ON THE PRESENCE OR ABSENCE OF A LEFT VENTRICULAR END DIASTOLIC DIAMETER (LVEDD) OF < 55MM. OF THE 1921 PATIENTS, 108 HAD LVEDD < 55MM, AND 1813 HAD LVEDD >= 55MM. RESULTS REPORTED THAT DEATH PRIOR TO HOSPITAL DISCHARGE WAS HIGHER IN THE SMALLER LVEDD GROUP (14.8 VS. 5.7%; P= 0.0007) AND SURVIVAL AT 2 YEARS WAS LOWER IN THE SMALLER LVEDD GROUP 63.3 VS. 81.8%; P= 0.0001).
IT WAS REPORTED THROUGH THE RESEARCH ARTICLE ¿LEFT VENTRICULAR DIMENSIONS AND CLINICAL OUTCOMES WITH A FULLY MAGNETICALLY LEVITATED LEFT VENTRICULAR ASSIST DEVICE¿ THAT HEARTMATE 3 (HM3) MAY BE ASSOCIATED WITH RIGHT HEART FAILURE, INFECTION, STROKE, BLEEDING, ARRYTHMIAS, MEDICAL/SURGICAL INTERVENTION, LOW FLOW, OTHER CARDIAC DISORDERS, NEUROLOGICAL DYSFUNCTION, THROMBOSIS, ANTI-COAGULATION RELATED ADVERSE EVENTS, RESPIRATORY FAILURE, RENAL DYSFUNCTION, HYPERTENSION RELATED ADVERSE EVENTS, SEPSIS, AND DEATH. THIS RETROSPECTIVE STUDY EVALUATED 1921 PATIENTS. THE STUDY AIMED TO EXAMINE THE ASSOCIATION OF SMALLER LEFT VENTRICULAR END-DIASTOLIC DIAMETER(LVEDD) WITH CLINICAL OUTCOMES. PATIENTS WITH MISSING LVEDD VALUES OR WITH ERRONEOUS BASELINE DATA WERE EXCLUDED FROM THE STUDY GROUP. THE PATIENT POPULATION WAS SPLIT BETWEEN THOSE WITH A SMALLER LVEDD (<55 MM) (N=108) AND THOSE WITH A LARGER LVEDD (>55MM) (N=1813) THE AVERAGE AGE OF PATIENTS WITH A SMALLER LVEDD WAS OLDER (63 +/- 11 YEARS) THAN THE AVERAGE AGE OF PATIENTS WITH A LARGER LVEDD (60 +/- 10 YEARS, P=0.005). PATIENTS WITH A SMALLER LVEDD WERE MORE OFTEN FEMALE (30.6%) COMPARED TO PATIENTS WITH A LARGER LVEDD (19.8%). OUT OF THE TOTAL 1921 EVALUATED PATIENTS, A TOTAL OF 338 PATIENTS PASSED AWAY. 66 PATIENTS PASSED AWAY DUE TO HEMOCOMPATIBILITY RELATED ADVERSE EVENTS. THE SITE CONSIDERED STROKE, BLEEDING, AND PUMP THROMBOSIS TO BE HEMOCOMPATIBILITY RELATED ADVERSE EVENTS, BUT THE BREAKDOWN BETWEEN THESE TYPES OF ADVERSE EVENTS WAS NOT REPORTED IN THE ARTICLE. 43 PATIENTS PASSED AWAY DUE TO MAJOR INFECTIONS, 130 PASSED AWAY DUE TO RIGHT HEART FAILURE. 119 PATIENTS PASSED AWAY PRIOR TO DISCHARGE AFTER IMPLANTATION. ADDITIONALLY, THE ARTICLE ALSO SITED NUMEROUS ADVERSE EVENTS EXPERIENCED BY THE PATIENT GROUP. THERE WERE 166 PATIENTS WHO EXPERIENCED A STROKE, OF WHICH 80 PATIENTS EXPERIENCED A HEMORRHAGIC STROKE, 96 PATIENTS EXPERIENCED AN ISCHEMIC STROKE, AND 79 PATIENTS EXPERIENCED A DEBILITATING STROKE. THERE WERE 666 PATIENTS WHO EXPERIENCED A CARDIAC ARRYTHMIA, OF WHICH 328 PATIENTS EXPERIENCED A SUPRAVENTRICULAR CARDIAC ARRYTHMIA, 374 PATIENTS EXPERIENCED A VENTRICULAR ARRYTHMIA, AND 37 EXPERIENCED BOTH. THERE WERE 932 PATIENTS WHO EXPERIENCED A MAJOR BLEEDING EVENT, OF WHICH 538 PATIENTS EXPERIENCED GASTROINTESTINAL BLEEDING, AND 243 REQUIRED SURGERY DUE TO THE BLEEDING. THERE WERE 1090 PATIENTS WHO EXPERIENCED AN INFECTION, OF WHICH 435 PATIENTS EXPERIENCED A DRIVELINE INFECTION, AND 287 PATIENTS EXPERIENCED SEPSIS. THERE WERE 108 PATIENTS WHO EXPERIENCED A HYPERTENSION ADVERSE EVENT. THERE WERE 289 PATIENTS WHO EXPERIENCED RENAL DYSFUNCTION. THERE WERE 384 PATIENTS WHO EXPERIENCED RESPIRATORY FAILURE THERE WERE 715 PATIENTS WHO EXPERIENCED RIGHT HEART FAILURE, OF WHICH 119 PATIENTS WERE TREATED WITH AN IMPLANTED RIGHT VENTRICLE ASSIST DEVICE (RVAD), AND 115 PATIENTS RECEIVED AN RVAD BEFORE INDEX DISCHARGE. THERE WERE ALSO 23 PATIENTS WITH SUSPECTED DEVICE THROMBOSIS. ADDITIONALLY, THE ARTICLE ALSO REPORTED NUMEROUS REASONS FOR REHOSPITALIZATIONS FOR PATIENTS WHILE ON THE HEARTMATE 3. IN TOTAL 1582 PATIENTS WERE HOSPITALIZED FOR ANY REASON. THERE WERE 517 REHOSPITALIZATIONS WITH BLEEDING AS THE PRIMARY CAUSE. THERE WERE 469 REHOSPITALIZATIONS WITH MEDICAL OR SURGICAL PROCEDURES AS THE PRIMARY CAUSE. THERE WERE 232 REHOSPITALIZATIONS WITH ANTICOAGULATION AS THE PRIMARY CAUSE. THERE WERE 566 REHOSPITALIZATIONS WITH MAJOR INFECTION AS THE PRIMARY CAUSE. THERE WERE 335 REHOSPITALIZATIONS WITH HEART FAILURE RELATED EVENTS AS THE PRIMARY CAUSE. THERE WERE 157 REHOSPITALIZATIONS WITH LOW FLOWS AS THE PRIMARY CAUSE. THERE WERE 248 REHOSPITALIZATIONS WITH CARDIAC ARRHYTHMIA AS THE PRIMARY CAUSE. THERE WERE 274 REHOSPITALIZATIONS WITH OTHER CARDIAC DISORDERS AS THE PRIMARY CAUSE. THERE WERE 152 REHOSPITALIZATIONS WITH NEUROLOGICAL DYSFUNCTION AS THE PRIMARY CAUSE. THERE WERE 274 REHOSPITALIZATIONS DUE TO OTHER UNSPECIFIED CARDIAC DISORDERS. THERE WERE 9 REHOSPITALIZATIONS DUE TO DEVICE THROMBOSIS. THERE WERE 790 REHOSPITALIZATIONS FOR OTHER UNSPECIFIED REASONS. FINALLY IT WAS REPORTED THAT PUMP EXCHANGE DUE TO DEVICE MALFUNCTION OCCURRED. THE STUDY CONCLUDED THAT SMALLER LVEDD WAS ASSOCIATED WITH INCREASED RISK FOR EARLY AND LATE MORTALITY, PRIMARILY AS A RESULT OF HEMOCOMPATIBILITY RELATED AND RIGHT HEART FAILURE RELATED DEATHS. IN ADDITION, RE-HOSPITALIZATIONS DUE TO LOW FLOWS WAS ALSO MORE COMMON IN SMALLER LVEDD PATIENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2590389 | HEARTMATE 3 LVAS IMPLANT KIT | VENTRICULAR (ASSIST) BYPASS | DSQ | THORATEC CORPORATION | 106524US | 00813024013297 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Death |