ARCHITECT IVANCOMYCIN
Report
- Report Number
- 3002809144-2010-00038
- Event Type
- Malfunction
- Date Received
- November 16, 2010
- Report Date
- October 16, 2010
- Manufacturer
- ABBOTT GERMANY
- Product Code
- LEH
- PMA / PMN Number
- K072036
- Removal / Correction Number
- 3002809144-9/15/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SA
- Reporter Occupation
- OTHER
Narratives
(B)(4). CONCLUSION: SAMPLE HANDLING IS BELIEVED TO BE A CONTRIBUTING FACTOR. ADDITIONAL REQUIREMENTS FOR CENTRIFUGATION HAVE ALSO BEEN ADDED TO THE HANDLING INSTRUCTIONS. THERE IS A POTENTIAL TO GENERATE FALSELY ELEVATED ARCHITECT I VANCOMYCIN PATIENT RESULTS DUE TO SAMPLE OR SAMPLE HANDLING ISSUES CAUSING INTERFERENCE. IN RESPONSE TO THE ISSUE A PRODUCT CORRECTION WAS INITIATED. ALL CURRENT CUSTOMERS WHO HAVE RECEIVED THE ARCHITECT IVANCOMYCIN ASSAY WERE SENT THE PRODUCT CORRECTION LETTER WITH SPECIFIC INSTRUCTIONS TO CONTINUE FOLLOWING ALL SAMPLE HANDLING INSTRUCTIONS PER THE PACKAGE INSERT AND TO FOLLOW THE ADDITIONAL CENTRIFUGATION INSTRUCTIONS INCLUDED IN THE PRODUCT CORRECTION LETTER. NEW CUSTOMERS WILL RECEIVE THE INFORMATION IN THE FORM OF A PRODUCT INFORMATION LETTER DISTRIBUTED WITH THE PRODUCT.
THE INVESTIGATION FOR FIELD ACTION FA13SEP2010 DETERMINED THE ROOT CAUSE OF ARCHITECT IVANCOMYCIN (LIST NUMBER 01P30-25) FOR FALSELY ELEVATED RESULTS IS DUE TO THE TRACER DILUENT FORMULATION. THE FORMULATION DOES NOT ADEQUATELY PROTECT AGAINST SPECIMEN CONTAMINATION. THE INVESTIGATION DETERMINED, AND SUBSEQUENT VERIFICATION CONFIRMED, A CHANGE TO THE CONJUGATE DILUENT WILL CORRECT THE ISSUE RELATED TO FALSELY ELEVATED RESULTS DUE TO CONTAMINATION, AND WILL ELIMINATE THE NEED FOR CUSTOMERS TO PERFORM ADDITIONAL SAMPLE CENTRIFUGATION AS INSTRUCTED IN PRODUCT CORRECTION LETTER OF FA13SEP2010. A PRODUCT INFORMATION LETTER REGARDING THIS NEW REAGENT WAS ISSUED ON (B)(4), 2012 INFORMING EX-US CUSTOMERS THAT THE NEW REAGENT FORMULATION, LIST NUMBER 1P30-27 WILL BE AVAILABLE (B)(4), 2012. THE ADDITIONAL SAMPLE CENTRIFUGATION AS INSTRUCTED IN PRODUCT CORRECTION LETTER OF FA13SEP2010 WILL CONTINUE TO BE FOLLOWED BY US CUSTOMERS. SUBMISSION FOR THE US 510K CLEARANCE OF THE NEW PRODUCT WILL OCCUR THE (B)(4) OF 2012.
THE CUSTOMER STATED FALSELY ELEVATED VANCOMYCIN RESULTS WERE GENERATED ON THE ARCHITECT I1000SR ANALYZER. THE CUSTOMER INDICATED THE PHYSICIANS HAD COMPLAINED ABOUT THE ERRATIC RESULTS AND IT WAS AFFECTING THEIR ABILITY TO ADJUST PATIENT DOSAGES. AN EVALUATION HAS DETERMINED THERE IS A POTENTIAL FOR THE ARCHITECT VANCOMYCIN ASSAY TO GENERATE FALSELY ELEVATED RESULTS DUE TO SAMPLE OR SAMPLE HANDLING ISSUES CAUSING INTERFERENCE. IN RESPONSE TO THIS ISSUE A PRODUCT CORRECTION LETTER WAS SENT TO THE CUSTOMER. THERE WAS NO NEGATIVE IMPACT TO PATIENT MANAGEMENT REPORTED.