ANEURX AAADVANTAGE STENT GRAFT SYSTEM -HYDRO
Report
- Report Number
- 2953200-2010-02083
- Event Type
- Injury
- Date Received
- October 28, 2010
- Date of Event
- September 29, 2010
- Report Date
- September 29, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL, RESULTS: (ARTERIAL/VESSEL OCCLUSION), (REVERSE CONE SHAPED AORTIC NECK; SEVERELY TORTUOUS VESSELS). EVAL, CONCLUSION: (REVERSE CONE SHAPED AORTIC NECK; SEVERELY TORTUOUS VESSELS), (PLACING STENT GRAFT HIGHER THAN INTENDED).
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 3.5 CM ABDOMINAL AORTIC ANEURYSM AND A 3.5 CM ILIAC ANEURYSM. THE AORTIC NECK WAS REVERSE CONE SHAPED, 21 MM DIAMETER AT THE RENAL ARTERIES AND 24 MM IN DIAMETER 2.5 MM DISTALLY. THE ILIAC VESSELS WERE SEVERELY TORTUOUS AND MILDLY CALCIFIED. IT WAS REPORTED THAT THE TALENT BIFURCATED STENT GRAFT (MFR REPORT # 2953200-2010-02082) WAS INADVERTENTLY PULLED DOWN DURING THE REMOVAL OF ITS DELIVERY CATHETER, RESULTING IN A PROXIMAL TYPE I ENDOLEAK AND KINKING OF THE IPSILATERAL LIMB. THE PHYSICIAN STATED THE DELIVERY SYSTEM HAD BECOME CAUGHT ON THE SEVERELY TORTUOUS VESSEL. THE PHYSICIAN ELECTED TO INTERVENE BY PLACING AN ANEURX AORTIC CUFF; HOWEVER, THE CUFF WAS PLACED HIGHER THEN INTENDED, PARTIALLY COVERING THE RENAL ARTERY. THE PHYSICIAN ELECTED TO PLACE A BARE METAL RENAL STENT, WHICH MADE THE RENAL ARTERY PATENT. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM -HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00348815 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |