FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 1900448 · Received October 28, 2010

Report

Report Number
1828100-2010-01928
Event Type
Malfunction
Date Received
October 28, 2010
Date of Event
October 5, 2010
Report Date
October 28, 2010
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION IN PROGRESS, BUT NOT YET CONCLUDED.

Description of Event or Problem · 1

DURING ROUTINE TESTING OF THE DEVICE AT THE SERVICE CENTER, THE SERVICE TECHNICIAN REPORTED ON VALVE #1 THE EXPECTED BLACK CONNECTOR WAS MISSING. INSTEAD UNIT HAD THE WIRES CONNECTED TO THE VALVE WITH SPADE CONNECTORS. THE SERVICE TECHNICIAN REPORTED THAT ALTHOUGH THE VALVE FUNCTIONED PROPERLY, IT DID NOT MEET SPECIFICATIONS. THE SERVICE TECH REPLACED THE VALVE AND CABLE ASSEMBLY AS PART OF THE REPAIR. SINCE THE EVENT OCCURRED AT THE SERVICE CENTER, THERE WAS NO PT INVOLVEMENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO TCM II COOLING AND HEATING SYSTEM HEATING COOLING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP 4415

Patients

Seq Age Sex Outcome Treatment
1