FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM

MDR report key: 1900330 · Received October 26, 2010

Report

Report Number
2242352-2010-03170
Event Type
Malfunction
Date Received
October 26, 2010
Date of Event
September 9, 2010
Report Date
September 27, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: A VISUAL INSPECTION REVEALED THAT THE DELIVERY DEVICE WAS RETURNED ALONE. THE SEAL AND THE LOADING DEVICE WERE NOT RETURNED. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THE DEVICE WAS BLOODY. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT, "THE SEAL DID NOT DEPLOY" COULD NOT BE CONFIRMED DUE TO INSUFFICIENT EVIDENCE. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III PROXIMAL SEAL DID NOT DEPLOY. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25012304

Patients

Seq Age Sex Outcome Treatment
1 NA