FDA Adverse Event Injury Summary report: N

UNK_I/A HANDPIECE

MDR report key: 19001470 · Received March 28, 2024

Report

Report Number
3012236936-2024-00985
Event Type
Injury
Date Received
March 28, 2024
Report Date
May 3, 2024
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
HQC
PMA / PMN Number
K981116
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THROUGH FOLLOW UP IT WAS REPORTED TO JNJ THAT THE ISSUE OF THE INFECTIONS HAD BEEN RESOLVED BY THE CUSTOMER (INTERNALLY) AND THAT THE CAUSE WAS IDENTIFIED TO BE A NEW EMPLOYEE THAT WAS NOT FOLLOWING PROPER ASEPTIC PROTOCOLS. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

CORRECTION: IN FOLLOW UP #1 IT WAS REPORTED THAT THE ISSUE OF THE INFECTIONS HAD BEEN RESOLVED BY THE CUSTOMER (INTERNALLY) AND THAT THE CAUSE WAS IDENTIFIED TO BE A NEW EMPLOYEE THAT WAS NOT FOLLOWING PROPER ASEPTIC PROTOCOLS. IT WAS FAILED TO MENTION THAT THE J&J DEVICE IS NO LONGER A SUSPECT PRODUCT AS IT DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT BASED ON THE INFORMATION THAT WAS RECEIVED FROM THE CUSTOMER. THEREFORE, THE REPORTABILITY OF THIS CASE WAS DOWNGRADED, AND NO ADDITIONAL INFORMATION WILL BE PROVIDED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A CASE OF INFECTION IN WHICH THE CUSTOMER HAS NOT BEEN ABLE TO ISOLATE THE SOURCE AND BELIEVE IT COULD BE THEIR SOLO IRRIGATION/ASPIRATION (I/A) HANDPIECE. PATIENT WAS PROVIDED MEDICAL TREATMENT FOR THE INFECTION AND IS EXPECTED TO HAVE A FULL RECOVERY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2241857 UNK_I/A HANDPIECE UNIT, PHACOFRAGMENTATION HQC JOHNSON & JOHNSON SURGICAL VISION, INC. UNK -I/A HANDPIECE

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention