FDA Adverse Event Malfunction Summary report: N

6 SHOOTER SAEED MULTI-BAND LIGATOR

MDR report key: 1900134 · Received October 22, 2010

Report

Report Number
1037905-2010-00532
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 23, 2010
Report Date
September 23, 2010
Manufacturer
COOK ENDOSCOPY
Product Code
MND
PMA / PMN Number
K944220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE INFO IN THIS REPORT WAS PROVIDED TO US BY THE COOK FACILITY IN (B)(4) ON BEHALF OF A MEDICAL FACILITY IN (B)(6). (B)(4). EVAL: WE COULD NOT CONFIRM THE REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED TO COOK ENDOSCOPY FOR EVAL. AFTER A REVIEW OF THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED, WE CAN REPORT NO DISCREPANCIES OR ANOMALIES WERE OBSERVED. WE WERE UNABLE TO CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN FINISHED GOODS INVENTORY. A REVIEW OF THE TWELVE MONTH COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. BASED ON THIS REVIEW, THE LIKELIHOOD OF THIS TYPE OF REPORT IS RARE. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVAL. THE INSTRUCTIONS FOR USE DIRECT THE USER TO ATTACH THE TRIGGER CORD TO THE HOOK ON THE END OF THE LOADING CATHETER, LEAVING APPROXIMATELY 2 CM OF TRIGGER CORD BETWEEN THE KNOT AND THE HOOK. IF THE KNOT AND THE HOOK ARE PLACED CLOSER THAN 2CM, THIS CAN CREATE RESISTANCE WHEN WITHDRAWING THE LOADING CATHETER THROUGH THE LIGATOR HANDLE AND/OR ACCESSORY CHANNEL OF THE ENDOSCOPE. IF ADDITIONAL PRESSURE IS APPLIED TO THE LOADING CATHETER WHEN RESISTANCE IS ENCOUNTERED, THIS COULD CONTRIBUTE TO SEPARATION OF THE BLUE HOOK FROM THE LOADING CATHETER. PRIOR TO DISTRIBUTION, ALL 6 SHOOTER SAEED MULTI-BAND LIGATORS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MFG PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. THE LIKELIHOOD OF THIS TYPE OF OCCURRENCE IS RARE. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

IN PREPARATION FOR A BAND LIGATION PROCEDURE, THE PHYSICIAN SELECTED A COOK ENDOSCOPY 6 SHOOTER SAEED MULTI-BAND LIGATOR. WHEN LOADING THE BARREL ONTO THE ENDOSCOPE (OUTSIDE THE PT), THE BLUE HOOK SEPARATED FROM THE LOADING CATHETER. THIS OCCURRED WHEN THE LOADING CATHETER WAS BEING USED TO PULL THE TRIGGER CORD THROUGH THE ACCESSORY CHANNEL OF THE ENDOSCOPE. A DIFFERENT LIGATOR WAS LOADED ONTO THE ENDOSCOPE TO PERFORM THE PROCEDURE. NO MEDICAL OR SURGICAL INTERVENTION WAS PERFORMED IN RESPONSE TO THIS OCCURRENCE. SEVERAL ATTEMPTS WERE MADE IN AN ATTEMPT TO COLLECT ADDITIONAL INFO REGARDING PT IMPACT AND PRODUCT USAGE. THE COMPLAINANT DID NOT SPECIFY IF THE PT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6 SHOOTER SAEED MULTI-BAND LIGATOR MND, LIGATOR, ESOPHAGEAL MND COOK ENDOSCOPY W2801850

Patients

Seq Age Sex Outcome Treatment
1 ENDOSCOPE (UNK TYPE)