HEARTSTRING PROXIMAL SEAL, 4.5 MM
Report
- Report Number
- 2242352-2010-02760
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- August 3, 2010
- Report Date
- August 16, 2010
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K022880
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE SEAL WAS COMPLETELY UNRAVELED EXCEPT FOR THE LAST WINDING, THE SEAL WAS SOMEWHAT TWISTED IN THE CENTER, AND THE SAFETY TAB WAS AT THE END OF THE SEAL. THE SEAL WAS OUTSIDE OF THE DELIVERY TUBE, AND WAS RECEIVED SEPARATE FROM THE SPRINGS AND THE BLUE STRING WHICH WAS CUT. THE COLLAR WAS NOT PRESENT, THE PLUNGER HAD BEEN DEPRESSED. THE DEVICE WAS VERY BLOODY. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "SEAL CRACKED DURING USE" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE HEARTSTRING II PROXIMAL SEAL CRACKED. A REPLACEMENT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL REPORTED NO PT EFFECTS. THE PRODUCT WAS RETURNED.