FDA Adverse Event Malfunction Summary report: N

HEARTSTRING PROXIMAL SEAL, 4.5 MM

MDR report key: 1900103 · Received October 22, 2010

Report

Report Number
2242352-2010-02760
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
August 3, 2010
Report Date
August 16, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K022880
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE SEAL WAS COMPLETELY UNRAVELED EXCEPT FOR THE LAST WINDING, THE SEAL WAS SOMEWHAT TWISTED IN THE CENTER, AND THE SAFETY TAB WAS AT THE END OF THE SEAL. THE SEAL WAS OUTSIDE OF THE DELIVERY TUBE, AND WAS RECEIVED SEPARATE FROM THE SPRINGS AND THE BLUE STRING WHICH WAS CUT. THE COLLAR WAS NOT PRESENT, THE PLUNGER HAD BEEN DEPRESSED. THE DEVICE WAS VERY BLOODY. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "SEAL CRACKED DURING USE" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE HEARTSTRING II PROXIMAL SEAL CRACKED. A REPLACEMENT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL REPORTED NO PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING PROXIMAL SEAL, 4.5 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HS-1045 25010602

Patients

Seq Age Sex Outcome Treatment
1 NA