FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1900050 · Received November 8, 2010

Report

Report Number
2531779-2010-02191
Event Type
Malfunction
Date Received
November 8, 2010
Date of Event
October 9, 2010
Report Date
October 9, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED THAT THE PUMP DID NOT RESPOND TO PRESSES OF THE AUDIO BOLUS BUTTON. THE AUDIO BOLUS BUTTON WAS FOUND TO BE MISALIGNED. THE PUMP WAS INTERMITTENTLY RESPONDING TO PRESSES OF ALL BUTTONS. THE "OK" BUTTON CONTACT WAS ALSO MISALIGNED AND ADHESIVE WAS FOUND UNDER ALL BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT REPORTED THAT THE PUMP WAS NOT RESPONDING TO KEYPAD BUTTON PRESSES. SHE WAS UNABLE TO CONFIRM THE VERIFY SCREEN AS THE "UP," "DOWN," AND "OK" BUTTONS REPORTEDLY DID NOT FUNCTION. THE PT ALSO REPORTED THAT THE KEYPAD WAS PEELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 29 YR