FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1900050
·
Received November 8, 2010
Report
- Report Number
- 2531779-2010-02191
- Event Type
- Malfunction
- Date Received
- November 8, 2010
- Date of Event
- October 9, 2010
- Report Date
- October 9, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED THAT THE PUMP DID NOT RESPOND TO PRESSES OF THE AUDIO BOLUS BUTTON. THE AUDIO BOLUS BUTTON WAS FOUND TO BE MISALIGNED. THE PUMP WAS INTERMITTENTLY RESPONDING TO PRESSES OF ALL BUTTONS. THE "OK" BUTTON CONTACT WAS ALSO MISALIGNED AND ADHESIVE WAS FOUND UNDER ALL BUTTON CONTACTS.
Description of Event or Problem · 1
THE PT REPORTED THAT THE PUMP WAS NOT RESPONDING TO KEYPAD BUTTON PRESSES. SHE WAS UNABLE TO CONFIRM THE VERIFY SCREEN AS THE "UP," "DOWN," AND "OK" BUTTONS REPORTEDLY DID NOT FUNCTION. THE PT ALSO REPORTED THAT THE KEYPAD WAS PEELING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |