FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 1899709 · Received November 5, 2010

Report

Report Number
2024168-2010-02359
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
October 8, 2010
Report Date
October 12, 2010
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
LOX
PMA / PMN Number
P810046
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE 3.0 X 18 MM XIENCE V (PART 1009541-18, LOT 0041241) IS BEING FILED UNDER A SEPARATE MFR #. EVAL SUMMARY: FACTORS THAT CAN CONTRIBUTE TO BALLOON RUPTURES INCLUDE, BUT ARE NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, WEAK MATERIALS, EXCESSIVELY APPLIED PRESSURE, OR INTERACTIONS WITH ACCESSORY DEVICES, PT ANATOMY, AND/OR LESION CALCIFICATION OR TORTUOSITY. RETURN OF THE VOYAGER NC CATHETER USED IN THE PROCEDURE MAY HAVE AIDED IN THE EVAL. THERE WAS NO REPORT OF ANY LEAK IN THE CATHETER NOTED DURING PREPARATION FOR USE, WHICH WOULD SUGGEST THAT THE BALLOON WAS NOT DAMAGED PRIOR TO USE. THE PT ANATOMY WAS MODERATELY TORTUOUS AND HEAVILY CALCIFIED AND THE VOYAGER NC WAS USED TO POST DILATE A STENT, WHICH LIKELY CONTRIBUTED TO THE REPORTED RUPTURE. IT IS POSSIBLE THAT THE BALLOON MATERIAL WAS DAMAGED (SCRATCHED) DURING INTERACTIONS WITH OTHER DEVICES, OR PT ANATOMY, SUCH THAT THE BALLOON RUPTURED UPON THE FIRST INFLATION AT 12 ATMS WHICH IS BELOW THE RATED BURST PRESSURE (RBP). A REVIEW OF THE PRODUCT MANUFACTURING RECORDS DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. THERE IS NO INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY. WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED BALLOON RUPTURE CANNOT BE DETERMINED. ALL DILATATION CATHETERS ARE 100% VISUALLY INSPECTED AND LEAK TESTED PRIOR TO PACKAGING. ADDITIONALLY, A SAMPLING OF UNITS IS DESTRUCTIVELY TESTED TO VERIFY RBP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE OF THE LEFT ANTERIOR DESCENDING ARTERY WHICH HAD MODERATE TORTUOSITY, HEAVY CALCIFICATION, ECCENTRIC DE NOVO, AND 90% STENOSIS, AFTER PRE-DILATATION WITH A BALLOON CATHETER AN "INDENTATION" WAS STILL NOTED BUT A 3.0 X 18 MM XIENCE V WAS IMPLANTED. THE BALLOON RUPTURED DURING THE FIRST INFLATION OF 30 SECONDS AT 10 ATMOSPHERES (ATMS). THE PHYSICIAN ATTEMPTED TO POST-DILATE THE 3.0 X 12 MM VOYAGER NC AT THE LESION, BUT IT ALSO RUPTURED DURING THE FIRST INFLATION OF 30 SECONDS AT 12 ATMS. THERE WAS NO REPORTED PT EFFECT. THE PHYSICIAN USED A NON-ABBOTT BALLOON TO POST-DILATE THE LESION AND THE PROCEDURE WAS COMPLETED WITHOUT ISSUE. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOYAGER NC CORONARY DILATATION CATHETER LOX ABBOTT VASCULAR-CARDIAC THERAPIES NA 0050362

Patients

Seq Age Sex Outcome Treatment
1 UNK FIELDER FC| GUIDE CATH: AXESS| GUIDE WIRE: RUNTHROUGH NS,| FORTIS| LOT 0041241)| DILATATION CATHETER: MAVERICK II,| STENT: 3.0 X 18 MM XIENCE V (PART 1009541-18,