FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 18993337 · Received March 28, 2024

Report

Report Number
9617229-2024-05381
Event Type
Injury
Date Received
March 28, 2024
Date of Event
February 14, 2024
Report Date
April 15, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE WAS RECEIVED ON APRIL 22, 2024, WITH LOT NUMBER 2905786. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED AN OPENING ON POSTERIOR ASSESSED AS A SURGICAL IMPACT OPENING. (SHELL THICKNESS WITHIN SPEC). ADDITIONAL OBSERVATIONS: NON-PENETRATING NICK OBSERVED ON THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED AS THE DEVICE WAS IMPLANTED.

Additional Manufacturer Narrative · 0

CORRECTION TO G.3. AWARE DATE OF SUPPLEMENTAL MEDWATCH #2. AWARE DATE SHOULD HAVE BEEN LISTED AS 21MAY2024.

Additional Manufacturer Narrative · 0

FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE, CONFIRMED VIA MAMMOGRAM. DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE, CONFIRMED VIA MAMMOGRAM. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1027833 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2905786

Patients

Seq Age Sex Outcome Treatment
1 74 YR Female Required Intervention