FDA Adverse Event Malfunction Summary report: N

ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM

MDR report key: 18992045 · Received March 28, 2024

Report

Report Number
3006575795-2024-00092
Event Type
Malfunction
Date Received
March 28, 2024
Date of Event
March 18, 2024
Report Date
March 27, 2024
Manufacturer
ZYNO MEDICAL, LLC
Product Code
FRN
UDI-DI
00814371020013
PMA / PMN Number
K130690
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ON 3/18/2024 FLOWRATE ISSUE WAS REPORTED WITHIN ZYNO MEDICAL, LLC WHICH WAS OBSERVED DURING CALIBARATION/ PREVENTIVE MAINTENANCE. THERE IS NO SPECIFIC CAUSE ESTABLISHED. A CAPA HAS BEEN OPENED IN ORDER TO FULLY DIAGNOSE AND ADDRESS THE ROOT CAUSE OF THE REPORTED EVENT. AS THIS IS OBSERVED DURING PREVENTIVE MAINTENANCE, ALL THE ISSSUES WITHIN THE PUMP WERE RESOLVED.

Description of Event or Problem · 0

ON 03/18/2024, ZYNO TECHNICIAN REPORTED AN ISSUE WITHIN ZYNO, LLC WHICH IS TAKEN AS AN COMPLAINT THAT THERE WAS A FLOW RATE ISSUE OBSERVED DURING PREVENTIVE MAINTENACE/ CALIBRATION WHERE FLOW RATE OFFSET% AT PRE-REWORK IS 6% AND FLOW RATE OFFET% AT POST-REWORK IS -1% . THIS IS DETERMINED TO BE A FLOWRATE UNKOWN ISSUE. NO PATIENT INVOLVED AS THIS IS OBSERVED DURING CALIBRATION/PREVENTIVE MAINTENANCE

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
719440 ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM LARGE VOLUME INFUSION PUMP FRN ZYNO MEDICAL, LLC Z800F 20160111-SH 00814371020013

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown