ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM
Report
- Report Number
- 3006575795-2024-00092
- Event Type
- Malfunction
- Date Received
- March 28, 2024
- Date of Event
- March 18, 2024
- Report Date
- March 27, 2024
- Manufacturer
- ZYNO MEDICAL, LLC
- Product Code
- FRN
- UDI-DI
- 00814371020013
- PMA / PMN Number
- K130690
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 003
Narratives
ON 3/18/2024 FLOWRATE ISSUE WAS REPORTED WITHIN ZYNO MEDICAL, LLC WHICH WAS OBSERVED DURING CALIBARATION/ PREVENTIVE MAINTENANCE. THERE IS NO SPECIFIC CAUSE ESTABLISHED. A CAPA HAS BEEN OPENED IN ORDER TO FULLY DIAGNOSE AND ADDRESS THE ROOT CAUSE OF THE REPORTED EVENT. AS THIS IS OBSERVED DURING PREVENTIVE MAINTENANCE, ALL THE ISSSUES WITHIN THE PUMP WERE RESOLVED.
ON 03/18/2024, ZYNO TECHNICIAN REPORTED AN ISSUE WITHIN ZYNO, LLC WHICH IS TAKEN AS AN COMPLAINT THAT THERE WAS A FLOW RATE ISSUE OBSERVED DURING PREVENTIVE MAINTENACE/ CALIBRATION WHERE FLOW RATE OFFSET% AT PRE-REWORK IS 6% AND FLOW RATE OFFET% AT POST-REWORK IS -1% . THIS IS DETERMINED TO BE A FLOWRATE UNKOWN ISSUE. NO PATIENT INVOLVED AS THIS IS OBSERVED DURING CALIBRATION/PREVENTIVE MAINTENANCE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 719440 | ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM | LARGE VOLUME INFUSION PUMP | FRN | ZYNO MEDICAL, LLC | Z800F | 20160111-SH | 00814371020013 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |