AUTOMATED IMPELLA CONTROLLER
Report
- Report Number
- 1220648-2024-08802
- Event Type
- Malfunction
- Date Received
- March 27, 2024
- Date of Event
- March 12, 2024
- Report Date
- April 19, 2024
- Manufacturer
- ABIOMED, INC
- Product Code
- OZD
- UDI-DI
- 00813502010985
- PMA / PMN Number
- P140003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE INVESTIGATION FOR THE REPORTED CONSOLE (AIC) PURGE ISSUES. THE IMPELLA WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE PURGE ASSEMBLY REVEALED EXCESSIVE PURGE FLUID RESIDUE, INDICATING LEAKAGE OF PURGE FLUID DURING CLINICAL USE. PURGE FLUID RESIDUE WAS ALSO FOUND WITHIN THE PURGE MOTOR ASSEMBLY, PREVENTING THE PURGE MOTOR FROM SPINNING BOTH BY HAND AND VIA TELNET COMMANDS TO THE CONSOLE. LOGS WERE REVIEWED WHICH CONFIRMED THE REPORTED COMPLAINT. THERE WAS AN INSTANCE OF A PURGE SYSTEM FAILURE ALARM DUE TO A BLOCKED PISTON. THE ROOT CAUSE OF THE PURGE SYSTEM FAILURE WAS DUE TO BUILDUP OF GLUCOSE ON THE PURGE STEPPER MOTOR GASKET.
THE IMPELLA DEVICE WAS RETURNED, AND AN EVALUATION IS UNDERWAY. UPON INVESTIGATION CLOSURE, A SUPPLEMENTAL MDR WILL BE FILED.
THE USER FACILITY REPORTED A 64-YEAR-OLD MALE IN ST-SEGMENT ELEVATION MYOCARDIAL INFARCTION WAS IMPLANTED WITH AN IMPELLA 5.5 DEVICE FOR MECHANICAL CIRCULATORY SUPPORT. IT WAS REPORTED THAT THE AUTOMATED IMPELLA CONTROLLER (AIC) PURGE SYSTEM FAILURE ALARM OCCURRED. IT WAS REPORTED THAT INSIDE THE PURGE DOOR WHERE THE CASSETTE FITS IN WHICH ROTATES FOR PURGE CASSETTE FUNCTION SEEMED TO BE LOCKED UP OR FROZEN ACCORDING TO CUSTOMER. THE AIC WAS EXCHANGED AND THERE WAS NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 494343 | AUTOMATED IMPELLA CONTROLLER | AUTOMATED IMPELLA CONTROLLER (AIC) | OZD | ABIOMED, INC | AUTOMATED IMPELLA CONTROLLER | 1427009 | 00813502010985 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Male |