FDA Adverse Event Malfunction Summary report: N

BARCO

MDR report key: 18985050 · Received March 27, 2024

Report

Report Number
3004455125-2024-00014
Event Type
Malfunction
Date Received
March 27, 2024
Date of Event
February 23, 2024
Report Date
March 27, 2024
Manufacturer
BARCO NV
Product Code
KQM
UDI-DI
05415334043312
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE UPDATED FIELDS AND FINAL INVESTIGATION RESULTS. THE CUSTOMER HAS PROVIDED ADDITIONAL INFORMATION FROM THEIR INVESTIGATION. THE DETACHMENT OF THE DISPLAY COVER WAS CAUSED BY INCORRECT USE OF THE ARM ON WHICH THE DISPLAY WAS MOUNTED. THE HANDLES PROVIDED FOR MOVING THE DISPLAY ON THE SPRING ARM WERE NOT USED. INSTEAD, THE MONITOR WAS GRIPPED AND MOVED BY THE RECES OR CABLE ENTRY. THE USERS WERE ADVISED BY THE CUSTOMER NOT TO DO THIS IN THE FUTURE. ADDITIONALLY, CABLE COVERS WERE SECURED WITH CABLE TIES. ADDITIONALLY THE CUSTOMER CONFIRMED THAT THE EVENT OCCURRED BEFORE SURGERY AND NOT DURING SURGERY. THERE WAS NO ADDITIONAL ANESTHESIA AND NO PROLONGATION OF THE PROCEDURE. THE PATIENT WAS NOT HARMED.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE UPDATED FIELDS AND FINAL INVESTIGATION RESULTS. THE CUSTOMER HAS PROVIDED ADDITIONAL INFORMATION FROM THEIR INVESTIGATION. THE DETACHMENT OF THE DISPLAY COVER WAS CAUSED BY INCORRECT USE OF THE ARM ON WHICH THE DISPLAY WAS MOUNTED. THE HANDLES PROVIDED FOR MOVING THE DISPLAY ON THE SPRING ARM WERE NOT USED. INSTEAD, THE MONITOR WAS GRIPPED AND MOVED BY THE RECES OR CABLE ENTRY. THE USERS WERE ADVISED BY THE CUSTOMER NOT TO DO THIS IN THE FUTURE. ADDITIONALLY, CABLE COVERS WERE SECURED WITH CABLE TIES. ADDITIONALLY THE CUSTOMER CONFIRMED THAT THE EVENT OCCURRED BEFORE SURGERY AND NOT DURING SURGERY. THERE WAS NO ADDIONAL ANESTHESIA AND NO PROLONGATION OF THE PROCEDURE. THE PATIENT WAS NOT HARMED.

Description of Event or Problem · 0

COVER FOR CABLE CONNECTIONS (PLASTIC COVER) FELL WHEN MOVING THE MONITOR ON THE OPERATING TABLE DURING SURGICAL PREPARATION. OPERATING TABLE UNSTERILE LONGER OPERATION AND ANESTHESIA TIME FOR THE PATIENT. COVER PROBABLY CAME LOOSE DURING CLEANING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
527123 BARCO SURGICAL DISPLAY KQM BARCO NV MDSC-8527 05415334043312

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown