FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 18984596 · Received March 27, 2024

Report

Report Number
9610877-2024-52520
Event Type
Malfunction
Date Received
March 27, 2024
Date of Event
March 15, 2024
Report Date
March 27, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
UDI-DI
04961333234066
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC34-I10TL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K180285. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WITH DRIVE PCB BLACKOUT. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE WITH DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE BRAND PLATE WORN OUT; HOWEVER, THIS DEFECT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
493274 PENTAX VIDEO COLONOSCOPE FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC34-I10TL 04961333234066

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown