FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 18983926 · Received March 26, 2024

Report

Report Number
3004753838-2024-071782
Event Type
Malfunction
Date Received
March 26, 2024
Date of Event
March 3, 2024
Report Date
April 1, 2024
Manufacturer
DEXCOM, INC.
Product Code
QBJ
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4) 3004753838-2024-071782 WAS REPORTED IN ERROR. PLEASE DISREGARD INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE.

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.

Description of Event or Problem · 0

THE COMPLAINT STATES THAT A TRANSMITTER BATTERY ISSUE WAS REPORTED. DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION WAS CONFIRMED VIA DATA AS A LOW TRANSMITTER BATTERY. THE PROBABLE CAUSE OF THE LOW BATTERY ALERT WAS DETERMINED TO BE A TRANSMITTER FAILURE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468787 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC.

Patients

Seq Age Sex Outcome Treatment
1 68 YR Female