FDA Adverse Event Malfunction Summary report: N

BLANKETROL II

MDR report key: 1898368 · Received October 28, 2010

Report

Report Number
1516825-2010-00012
Event Type
Malfunction
Date Received
October 28, 2010
Date of Event
September 30, 2010
Report Date
September 30, 2010
Manufacturer
CINCINNATI SUB-ZERO PRODUCTS, INC.
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER IS NOT RETURNING THE UNIT. ENGINEERING HAS EVALUATED THE INFO AND NO CONCLUSIONS CAN BE DRAWN. IN ORDER TO FURTHER INVESTIGATE, CSZ HAS REQUESTED ADDITIONAL INFO TO IDENTIFY ROOT CAUSE OF THE PROBLEM. A SUPPLEMENTAL REPORT WILL BE SUBMITTED TO FDA AS SOON AS NEW INFO OR INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 1

THE DISTRIBUTOR ALLEGED, "IN THE HEATING MODE WHEN WE SET TEMP TO 37C, HI-LIMIT ALARM ACTIVATES WHEN SYSTEM REACHES TO 33C AND MACHINE SHUTS OFF. WHEN AFTER SOME TIME, RESTART THE MACHINE WITH THE SAME TEMP I.E 37C. REACHING 37C, MACHINE CHANGES ITS MODE TO COOLING, BUT TEMP RISES TO 42C. AT THIS TIME, THE MACHINE IS DISPLAYING THE COOLING MODE BUT IS WORKING IN THE HEATING MODE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BLANKETROL II THERMAL REGULATING SYSTEM DWJ CINCINNATI SUB-ZERO PRODUCTS, INC. 222R

Patients

Seq Age Sex Outcome Treatment
1