FDA Adverse Event Injury Summary report: N

VIVA¿ XT CRT-D

MDR report key: 18981949 · Received March 26, 2024

Report

Report Number
3004209178-2024-07864
Event Type
Injury
Date Received
March 26, 2024
Date of Event
February 13, 2024
Report Date
April 16, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
NIK
UDI-DI
00643169530140
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS FURTHER REPORTED THAT THE PATIENT RECEIVED RADIOFREQUENCY KYPHOPLASTY (RFK) WHICH WAS IDENTIFIED AS THE CAUSE OF THE CONFIRMED ELECTROMAGNETIC INTERFERENCE WHICH LEAD TO THE INAPPROPRIATE SHOCKS.

Description of Event or Problem · 0

IT WAS REPORTED THAT  THE PATIENT EXPERIENCED A POTENTIALLY INAPPROPRIATE SHOCK. UPON REVIEW IT WAS NOTED THAT THERE WAS RIGHT ATRIA L (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD OVERSENSING, RA LEAD UNDERSENSING AND POTENTIAL ELECTROMAGNETIC INTERFERENCE FROM THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D).  THE RA LEAD, RV LEAD AND CRT-D REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
441465 VIVA¿ XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC PUERTO RICO OPERATIONS CO. DTBA1D1 00643169530140

Patients

Seq Age Sex Outcome Treatment
1 76 YR Female Life Threatening 5076-52 LEAD, 419488 LEAD, 6944-65 LEAD