ATTAIN STABILITY QUAD MRI SURESCAN

MDR report key: 18980159 · Received March 26, 2024

Report

Report Number
2649622-2024-08249
Event Type
Malfunction
Date Received
March 26, 2024
Date of Event
February 28, 2024
Report Date
April 16, 2024
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE ANALYST NOTED THE GUIDEWIRE WAS NOT RETURNED. THE SAMPLE OF ATTAIN GWR 419488 WAS USED TO PERFORM TEST. BY BACK LOADING, IT WAS PASSED THE COIL LUMEN WITHOUT ISSUE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE GUIDEWIRE STOPPED ENTERING THE LEFT VENTRICULAR (LV) LEAD AND IT WAS SUSPECTED THAT THE BLOOD ON THE GUIDEWIRE BECAME STUCK INSIDE THE LEAD. A SECOND LV LEAD WAS PLACED, HOWEVER, THE GUIDEWIRE DID NOT ENTER THE LEAD TIP AND A DEFECT WITH THE LV LEAD WAS SUSPECTED. THE LV LEADS WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
14743 ATTAIN STABILITY QUAD MRI SURESCAN DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P OJX MPRI 479888

Patients

Seq Age Sex Outcome Treatment
1 68 YR Male