FDA Adverse Event Malfunction Summary report: N

ADVANTA2 RENTAL BED

MDR report key: 18977778 · Received March 26, 2024

Report

Report Number
3006697241-2024-00023
Event Type
Malfunction
Date Received
March 26, 2024
Date of Event
February 27, 2024
Report Date
March 26, 2024
Manufacturer
HILL-ROM MEXICO
Product Code
FNJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE BAXTER TECHNICIAN FOUND THE CASTERS NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. CHECK THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE BAR OR THE LH/RH BRAKE PEDALS ARE PRESSED. REPAIR AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN SEPTEMBER 2023. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE CASTERS TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 0

BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
499921 ADVANTA2 RENTAL BED BED, MANUAL FNJ HILL-ROM MEXICO P1190ARENT01

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown