Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
MDR report key: 1897114
·
Received November 11, 2010
Report
- Report Number
- 2050012-2010-01250
- Event Type
- Malfunction
- Date Received
- November 11, 2010
- Date of Event
- October 25, 2010
- Report Date
- November 10, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE AND REMOVED AND REPLACED LEVEL SENSE BEAD.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REPORTING THAT THE REAGENT PROBE COLLAR WASH WAS SPITTING ON THE INSTRUMENT. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |