FDA Adverse Event Malfunction Summary report: N

UROLOGICAL BALLOON DILATATION CATHETER

MDR report key: 1897019 · Received November 11, 2010

Report

Report Number
3005099803-2010-04629
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
October 15, 2010
Report Date
October 21, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
EZN
PMA / PMN Number
K980795
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A VISUAL AND TACTILE EVALUATION OF THE RETURNED DEVICE REVEALED NO DAMAGE TO THE SHAFT OF THE DEVICE. A FUNCTIONAL EVALUATION OF THE DEVICE REVEALED A LEAK IN THE BALLOON WHEN IT WAS ATTACHED TO AN ENCORE INFLATION UNIT AND WAS INFLATED TO ITS RATED BURST PRESSURE (RBP). MICROSCOPIC EVALUATION OF THE BALLOON REVEALED A LONGITUDINAL TEAR 6MM PROXIMAL TO THE TIP OF THE DEVICE AND WHICH EXTENDED 42MM PROXIMALLY. SEVERAL SCRATCHES WERE ALSO NOTED ON THE BALLOON MATERIAL.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA BALLOON DILATATION CATHETER WAS USED DURING AN UNKNOWN PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON RUPTURED DURING INFLATION. THE BALLOON WAS INFLATED WITH 4 CC OF CONTRAST SOLUTION AT THE TIME OF THE FAILURE. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "FINE".

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA BALLOON DILATATION CATHETER WAS USED DURING AN UNKNOWN PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON RUPTURED DURING INFLATION. THE BALLOON WAS INFLATED WITH 4 CC OF CONTRAST SOLUTION AT THE TIME OF THE FAILURE. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "FINE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UROLOGICAL BALLOON DILATATION CATHETER DILATOR, CATHETER, URETERAL EZN BOSTON SCIENTIFIC - GALWAY M0062251200 13731513

Patients

Seq Age Sex Outcome Treatment
1