FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1896997 · Received November 11, 2010

Report

Report Number
1423500-2010-05648
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
October 3, 2010
Report Date
October 19, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE EVALUATION DID NOT CONFIRM ANY FAILURE OR MALFUNCTION OF THE DEVICE THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS DISCOVERED DURING EVALUATION. BASED ON A REVIEW OF ALL AVAILABLE THERAPY LOG DATA, THE CAUSE OF THE IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN, / USE ERROR AS THE TIDAL ULTRAFILTRATION REMOVAL WAS SET TOO LOW (50 ML). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE IIPV. A LABELING REVIEW FOUND LABELING TO BE ADEQUATE FOR THE USE ERROR IDENTIFIED IN THIS REPORT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 9. THE PATIENT'S ULTRAFILTRATION WAS 924ML. THE FILL VOLUME WAS 1100ML. THIS MEETS IIPV CRITERIA. PRODUCT SURVEILLANCE CONTACTED THE NURSE ON (B)(6) 2010. ACCORDING TO THE NURSE SHE THE PATIENT'S TDTAL ULTRAFILTRATION REMOVAL IS SET LOW. THE NURSE STATED THAT SETTING THE ULTRAFILTRATION AT 80% IS THE ONLY WAY FOR THE PATIENT TO MANAGE HER FLUIDS. THE NURSE STATED THEY ARE MONITORING THE PATIENT AND HER SETTINGS. THE PATIENT HAS NOT REPORTED ANY SYMPTOMS OF OVERFILL. THE PATIENT HAS RESUMED THERAPY SUCCESSFULLY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 40 YR