FDA Adverse Event Malfunction Summary report: N

COLLEAGUE COLOR PUMP CE FRENCHVERSION SINGLE

MDR report key: 1896974 · Received November 11, 2010

Report

Report Number
6000001-2010-04808
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
August 31, 2010
Report Date
September 20, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BATTERY ISSUE WAS CONFIRMED BY BAXTER PERSONNEL AS A DEPLETED BATTERY ALARM. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES. THIS CONDITION WAS RESOLVED AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED A BATTERY ISSUE. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. UPON REVIEWING THE PUMP'S EVENT HISTORY, BAXTER QUALITY ENGINEERING DISCOVERED THIS EVENT INTERRUPTED DELIVERY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.63.90.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE COLOR PUMP CE FRENCHVERSION SINGLE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1