SCALPEL HOLDER RED; ASPEN LEAF WITH BARD-PARKER; STERILE
Report
- Report Number
- 3002750084-2024-00010
- Event Type
- Malfunction
- Date Received
- March 25, 2024
- Report Date
- July 15, 2024
- Manufacturer
- ASPEN SURGICAL PRODUCTS, INC.
- Product Code
- FSM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT PRODUCT WAS FOUND WITH SEAL ISSUES. THE DEVICE WAS RETURNED FOR EVALUATION AND THE SEAL ISSUES WERE CONFIRMED. IT WAS DETERMINED THAT THE SCALPEL TRAY IN THE POUCH CONTAINED PART OF THE SPLICE FROM THE TYVEK ROLL THAT ASPEN RECEIVES FROM THE VENDOR. IT WAS MISSED DURING ASPEN'S PRODUCTION RUN BECAUSE THE LABEL FACES UP TO THE OPERATOR AND THE RED SPLICE WAS UNDETRECTED SINCE IT FACES DOWNWARD. AWARENESS TRAINING WAS EXECUTED. THIS CAN BE CONSIDERED THE FINAL REPORT, HOWEVER IF ASPEN IDENTIFIES ANY ADDITIONAL RELEVANT INFORMATION IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
ASPEN SURGICAL RECIEVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT PRODUCT WAS FOUND WITH SEAL ISSUES. THE ACTUAL DEVICE WILL BE RETURNED FOR EVALUATION. THE MANUFACTURING LOT NUMBERS WERE PROVIDED FOR REVIEW. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED, THE ADDITIONAL RELEVANT INFORMATION WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
ASPEN SURGICAL RECEIVED A REPORT FROM OUTSIDE DISTRIBUTOR INDICATING THAT PRODUCT ST-100 WAS DISCOVERED WITH SEALING ISSUES. NO INJURY/DEATH WAS REPORTED. THIS IS ENTERED IN OUR SYSTEM AS (B)(4).
ASPEN SURGICAL RECEIVED A REPORT FROM OUTSIDE DISTRIBUTOR INDICATING THAT PRODUCT ST-100 WAS DISCOVERED WITH SEALING ISSUES. NO INJURY/DEATH WAS REPORTED. THIS IS ENTERED IN OUR SYSTEM AS (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1724 | SCALPEL HOLDER RED; ASPEN LEAF WITH BARD-PARKER; STERILE | SCALPEL HOLDER | FSM | ASPEN SURGICAL PRODUCTS, INC. | ST-100 | 359853 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |