FDA Adverse Event Malfunction Summary report: N

CONNECTED BLOOD GLUCOSE MONITORING SYSTEM

MDR report key: 18969538 · Received March 25, 2024

Report

Report Number
3008514395-2024-00005
Event Type
Malfunction
Date Received
March 25, 2024
Date of Event
February 27, 2024
Report Date
March 25, 2024
Manufacturer
TELCARE LLC
Product Code
NBW
UDI-DI
00859519002407
PMA / PMN Number
K110571
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INSPECTION OF THE RETURNED MATERIALS CONFIRMED THAT THE PLASTIC AND GLUE ON THE USB-C CONNECTOR OF THE CHARGER MELTED INTO THE DEVICE CHARGING PORT CAUSING IT TO BECOME STUCK. APART FROM MELTED MATERIAL IN THE CHARGING PORT, THE RETURNED METER AND BATTERY SHOWED NO OTHER EVIDENCE OF DAMAGE OR MALFUNCTION. THE CAUSE OF THE ALLEGED MELTING WAS FOUND TO BE OVERHEATING OF THE USB-C CONNECTOR. PATIENT WAS ISSUED A REPLACEMENT DEVICE.

Description of Event or Problem · 0

PATIENT ALLEGED THAT THEIR BLOOD GLUCOSE METER CHARGER MELTED INTO THE METER ITSELF WHILE CONNECTED TO THE DEVICE. NO INJURY OR PROPERTY DAMAGE WAS ALLEGED AND NO MEDICAL ATTENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10746 CONNECTED BLOOD GLUCOSE MONITORING SYSTEM CONNECTED BLOOD GLUCOSE METER NBW TELCARE LLC TM0009 00859519002407

Patients

Seq Age Sex Outcome Treatment
1 69 YR Unknown