FDA Adverse Event
Malfunction
Summary report: N
CONNECTED BLOOD GLUCOSE MONITORING SYSTEM
MDR report key: 18969538
·
Received March 25, 2024
Report
- Report Number
- 3008514395-2024-00005
- Event Type
- Malfunction
- Date Received
- March 25, 2024
- Date of Event
- February 27, 2024
- Report Date
- March 25, 2024
- Manufacturer
- TELCARE LLC
- Product Code
- NBW
- UDI-DI
- 00859519002407
- PMA / PMN Number
- K110571
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
INSPECTION OF THE RETURNED MATERIALS CONFIRMED THAT THE PLASTIC AND GLUE ON THE USB-C CONNECTOR OF THE CHARGER MELTED INTO THE DEVICE CHARGING PORT CAUSING IT TO BECOME STUCK. APART FROM MELTED MATERIAL IN THE CHARGING PORT, THE RETURNED METER AND BATTERY SHOWED NO OTHER EVIDENCE OF DAMAGE OR MALFUNCTION. THE CAUSE OF THE ALLEGED MELTING WAS FOUND TO BE OVERHEATING OF THE USB-C CONNECTOR. PATIENT WAS ISSUED A REPLACEMENT DEVICE.
Description of Event or Problem · 0
PATIENT ALLEGED THAT THEIR BLOOD GLUCOSE METER CHARGER MELTED INTO THE METER ITSELF WHILE CONNECTED TO THE DEVICE. NO INJURY OR PROPERTY DAMAGE WAS ALLEGED AND NO MEDICAL ATTENTION WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10746 | CONNECTED BLOOD GLUCOSE MONITORING SYSTEM | CONNECTED BLOOD GLUCOSE METER | NBW | TELCARE LLC | TM0009 | 00859519002407 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Unknown |