ECHELON 45
Report
- Report Number
- 3005075853-2010-06441
- Event Type
- Malfunction
- Date Received
- November 11, 2010
- Date of Event
- October 14, 2010
- Report Date
- October 14, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051002
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4): PAN. THE ANALYSIS RESULTS THAT ONE (B)(4) DEVICE WAS RETURNED WITH NO VISUAL NON-CONFORMANCES AND A RELOAD PAN WAS FOUND LODGE INSIDE THE CHANNEL. THERE WAS NO CARTRIDGE RELOAD PRESENT. THE PAN WAS REMOVED FROM THE CHANNEL AND THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD. THE DEVICE FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTE TO HAVE THE PROPER B-FORMED SHAPE. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE PAN BECAME DISLODGE FROM THE RELOAD, IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HEPATECTOMY PROCEDURE, THE DEVICE WAS GOING TO BE USED FOR THE SECOND FIRING. THE DEVICE CAME OFF FROM THE INSTRUMENT BEFORE FIRING; THE DEVICE COULD NOT BE USED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON 45 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) - RELOAD |