FDA Adverse Event Malfunction Summary report: N

ECHELON 45

MDR report key: 1896913 · Received November 11, 2010

Report

Report Number
3005075853-2010-06441
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
October 14, 2010
Report Date
October 14, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PAN. THE ANALYSIS RESULTS THAT ONE (B)(4) DEVICE WAS RETURNED WITH NO VISUAL NON-CONFORMANCES AND A RELOAD PAN WAS FOUND LODGE INSIDE THE CHANNEL. THERE WAS NO CARTRIDGE RELOAD PRESENT. THE PAN WAS REMOVED FROM THE CHANNEL AND THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD. THE DEVICE FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTE TO HAVE THE PROPER B-FORMED SHAPE. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE PAN BECAME DISLODGE FROM THE RELOAD, IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HEPATECTOMY PROCEDURE, THE DEVICE WAS GOING TO BE USED FOR THE SECOND FIRING. THE DEVICE CAME OFF FROM THE INSTRUMENT BEFORE FIRING; THE DEVICE COULD NOT BE USED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON 45 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1 (B)(4) - RELOAD