FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 1896613 · Received November 10, 2010

Report

Report Number
2017865-2010-05061
Event Type
Injury
Date Received
November 10, 2010
Date of Event
September 3, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD EXHIBITED LESS THAN 100 OHMS IMPEDANCE AND 0.2 MV P-WAVE SENSING. AN X-RAY REVEALED CLAVICULAR CRUSH DAMAGE TO THE INSULATION. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1882TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention