FDA Adverse Event
Injury
Summary report: N
TRILOGY DR+
MDR report key: 1896434
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-05244
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- August 17, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DXY
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
FINAL ANALYSIS FOUND THE PULSE GENERATOR TO BE IN BACK-UP VVI MODE DUE TO DEVICE BATTERY VOLTAGE BELOW END-OF-LIFE (EOL) LEVEL. THIS WAS DUE TO NORMAL BATTERY DEPLETION. AFTER REPLACING THE BATTERY AND DOWNLOADING THE PRODUCT CODE, NORMAL FUNCTION ENSUED.
Description of Event or Problem · 1
CALLER REPORTS LANCET PROTRUDES BEYOND THE END CAP OF THE SOFTCLIX DEVICE. NO ACCIDENTAL NEEDLE STICK OCCURRED. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND REPLACEMENT WAS SENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PULSE GENERATOR WENT ABRUPTLY FROM ELECTIVE REPLACEMENT INDICATOR (ERI) TO END OF LIFE (EOL). THE DEVICE WAS REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRILOGY DR+ | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 2360L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention |