FDA Adverse Event Injury Summary report: N

TRILOGY DR+

MDR report key: 1896434 · Received November 10, 2010

Report

Report Number
2017865-2010-05244
Event Type
Injury
Date Received
November 10, 2010
Date of Event
August 17, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DXY
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THE PULSE GENERATOR TO BE IN BACK-UP VVI MODE DUE TO DEVICE BATTERY VOLTAGE BELOW END-OF-LIFE (EOL) LEVEL. THIS WAS DUE TO NORMAL BATTERY DEPLETION. AFTER REPLACING THE BATTERY AND DOWNLOADING THE PRODUCT CODE, NORMAL FUNCTION ENSUED.

Description of Event or Problem · 1

CALLER REPORTS LANCET PROTRUDES BEYOND THE END CAP OF THE SOFTCLIX DEVICE. NO ACCIDENTAL NEEDLE STICK OCCURRED. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND REPLACEMENT WAS SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PULSE GENERATOR WENT ABRUPTLY FROM ELECTIVE REPLACEMENT INDICATOR (ERI) TO END OF LIFE (EOL). THE DEVICE WAS REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRILOGY DR+ IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 2360L

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention