ZEPHYR XL DR
Report
- Report Number
- 2017865-2010-05344
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- August 17, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DXY
- PMA / PMN Number
- P880086
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
FINAL ANALYSIS FOUND THE PULSE GENERATOR TO BE IN BACK-UP MODE DUE TO DEVICE BATTERY VOLTAGE DROPPING BELOW END-OF-LIFE (EOL) LEVEL. THIS WAS DUE TO NORMAL BATTERY DEPLETION. AFTER REPLACING THE BATTERY, NORMAL FUNCTION ENSUED.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED . DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
IT WAS REPORTED THAT THE SYMPTOMATIC PATIENT PRESENTED TO THE EMERGENCY ROOM IN BACKUP VVI MODE. THE PATIENT ALSO EXPERIENCE SYNCOPE AND WAS LOST TO FOLLOW-UP WITH NO PRIOR INTERROGATIONS ON THE DEVICE. THERE WAS NO PACING SUPPORT UNTIL DOWNLOAD ATTEMPT. DUE TO LOW BATTERY STATUS FURTHER TROUBLESHOOTING COULD NOT BE PERFORMED. THE PULSE GENERATOR WAS REPLACED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI).
OLYMPUS RECEIVED A MEDWATCH WHICH STATED "OLYMPUS LIGHT SOURCE FOR EGD (ESOPHAGO-GASTRODUODENOSTOMY) TUBE PLACEMENT WITH MD. BULB BURNED OUT - SWITCHED TO SPARE BULB ON MACHINE WHICH IS NOT ADEQUATE LIGHT FOR A PROCEDURE. SWITCHED TO ALTERNATE LIGHT SOURCE WHICH ITSELF HAS NOISE AND LOW LIGHT. PHYSICIAN WAS UNABLE TO COMPLETE PROCEDURE WITH EITHER PIECE OF EQUIPMENT. PROCEDURE TERMINATED. ALL THIS OCCURRED DURING THE PROCEDURE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZEPHYR XL DR | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 5826 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention |