FDA Adverse Event Injury Summary report: N

OPTISENSE

MDR report key: 1896353 · Received November 10, 2010

Report

Report Number
2017865-2010-05211
Event Type
Injury
Date Received
November 10, 2010
Date of Event
August 18, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPTISENSE PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1999/52 NA

Patients

Seq Age Sex Outcome Treatment
1 88 YR Required Intervention (B)(4)