FDA Adverse Event
Malfunction
Summary report: N
TENDRIL ST
MDR report key: 1896337
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-05161
- Event Type
- Malfunction
- Date Received
- November 10, 2010
- Date of Event
- August 13, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN, (B)(4) NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND AN INSULATION BREACH AT 4.5 CM FROM THE CONNECTOR PIN. LEAD CONTINUITY WAS NORMAL.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |