FDA Adverse Event
Malfunction
Summary report: N
TENDRIL ST
MDR report key: 1896331
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-05142
- Event Type
- Malfunction
- Date Received
- November 10, 2010
- Date of Event
- September 17, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN, (B)(4) FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND MEDICAL ADHESIVE ON THE RING ELECTRODE, CAUSING THE LEAD TO EXHIBIT HIGH IMPEDANCE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |