FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 1896310 · Received November 10, 2010

Report

Report Number
2017865-2010-05069
Event Type
Injury
Date Received
November 10, 2010
Date of Event
August 13, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD DISLODGED. A CHEST X-RAY REVEALED THAT THE LEAD HAD 16 TWISTS DUE TO TWIDDLER'S SYNDROME. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/46 NA

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention (B)(4)