FDA Adverse Event
Malfunction
Summary report: N
TENDRIL ST
MDR report key: 1896274
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-05114
- Event Type
- Malfunction
- Date Received
- November 10, 2010
- Date of Event
- August 9, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(6). FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND A PIECE OF COIL INSIDE THE CONNECTOR PORTION OF THE LEAD. THIS CAUSED A SHORT CIRCUIT BETWEEN THE COILS, WHICH LED TO LOSS OF CAPTURE AND SENSING. OTHER: NA.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |