FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4

MDR report key: 18961819 · Received March 22, 2024

Report

Report Number
1045254-2024-00405
Event Type
Malfunction
Date Received
March 22, 2024
Date of Event
February 23, 2024
Report Date
March 23, 2024
Manufacturer
MEDTRONIC XOMED INC.
Product Code
ERL
UDI-DI
00643169350687
PMA / PMN Number
K041413
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G3: CORRECT 510K IS K041413. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS COULD NOT VERIFY EXCESSIVE HEAT. BECAUSE OF CORRSION ON REAR SEALS, BEARINGS AND SHAFT NOT RUNNING. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT INTRA OP DEVICE WAS OVERHEATING. THERE IS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
523806 IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4 DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I ERL MEDTRONIC XOMED INC. 1898200T 211435655 00643169350687

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown