CURRENT RF DR
MDR report key: 1896041
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-04382
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- September 27, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED COMMUNICATION ANOMALY WAS CONFIRMED IN THE LABORATORY. BASED ON AVAILABLE PARAMETER AND USAGE INFORMATION, THE DEVICE WAS FOUND TO BE BELOW EXPECTED LIMITS. NO SOURCES OF HIGH CURRENT DRAIN WERE FOUND THROUGHOUT TESTING. THE ORIGINAL BATTERY WAS SENT TO THE VENDOR FOR FURTHER ANALYSIS. NO ANOMALIES WERE FOUND THAT WOULD CAUSE BATTERY DEPLETION. THE CAUSE OF THE DEPLETION WAS NOT DETERMINED.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CURRENT RF DR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 2207-36 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |