FDA Adverse Event Injury Summary report: N

CURRENT RF DR

MDR report key: 1896041 · Received November 10, 2010

Report

Report Number
2017865-2010-04382
Event Type
Injury
Date Received
November 10, 2010
Date of Event
September 27, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED COMMUNICATION ANOMALY WAS CONFIRMED IN THE LABORATORY. BASED ON AVAILABLE PARAMETER AND USAGE INFORMATION, THE DEVICE WAS FOUND TO BE BELOW EXPECTED LIMITS. NO SOURCES OF HIGH CURRENT DRAIN WERE FOUND THROUGHOUT TESTING. THE ORIGINAL BATTERY WAS SENT TO THE VENDOR FOR FURTHER ANALYSIS. NO ANOMALIES WERE FOUND THAT WOULD CAUSE BATTERY DEPLETION. THE CAUSE OF THE DEPLETION WAS NOT DETERMINED.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CURRENT RF DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 2207-36 NA

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention