FDA Adverse Event Injury Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 1895953 · Received November 10, 2010

Report

Report Number
2017865-2010-04428
Event Type
Injury
Date Received
November 10, 2010
Date of Event
September 21, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT SOME EPISODES WITH NON-SUSTAINED NOISE WERE OBSERVED. HIGH IMPEDANCE AND HIGH THRESHOLDS WERE NOTED. THE DECISION WAS MADE TO MONITOR THE PATIENT. THE LEAD REMAINS IMPLANTED.

Description of Event or Problem · 1

NEW INFORMATION RECEIVED NOTES INSULATION BREAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7002/60 NA

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention