FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM PASSIVE FIXATION

MDR report key: 1895917 · Received November 10, 2010

Report

Report Number
2017865-2010-04654
Event Type
Injury
Date Received
November 10, 2010
Date of Event
September 13, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED A LEAD TIP STIFFNESS TEST WAS PERFORMED AND WAS FOUND TO BE WITHIN SPECIFICATION.

Description of Event or Problem · 1

DURING A FOLLOW UP, A CAPTURE ANOMALY, LOW SENSING AND HIGH LEAD IMPEDANCE WERE OBSERVED. FLUOROSCOPY REVEALED THAT THE LEAD SLIGHTLY ADVANCED INTO THE RIGHT VENTRICLE. THE PHYSICIAN SUSPECTED A PERICARDIAL EFFUSION. HENCE, THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7170/65 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention