- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED A LEAD TIP STIFFNESS TEST WAS PERFORMED AND WAS FOUND TO BE WITHIN SPECIFICATION.
Description of Event or Problem · 1
DURING A FOLLOW UP, A CAPTURE ANOMALY, LOW SENSING AND HIGH LEAD IMPEDANCE WERE OBSERVED. FLUOROSCOPY REVEALED THAT THE LEAD SLIGHTLY ADVANCED INTO THE RIGHT VENTRICLE. THE PHYSICIAN SUSPECTED A PERICARDIAL EFFUSION. HENCE, THE LEAD WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM PASSIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7170/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |