FDA Adverse Event
Injury
Summary report: N
RIATA ST OPTIM ACTIVE FIXATION
MDR report key: 1895915
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-04442
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- September 13, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY FOR NOISE ON THE VENTRICULAR LEAD. A MESSAGE WAS DISPLAYED THAT THERE WAS POSSIBLE OUTPUT CIRCUIT DAMAGE. THE IMPEDANCE WAS 0 OHMS. THE ICD AND LEAD WERE EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ST OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7021/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention | 1207-36, (B)(4) |