FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, SJ4 CONNECTOR
MDR report key: 1895908
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-04554
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- September 1, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P950022
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED OTHER TEXT : DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.
Description of Event or Problem · 1
IT WAS LATER REPORTED THAT THE LEAD WAS EXPLANTED WHEN REPOSITIONING WAS UNSUCCESSFUL DUE TO DISLODGEMENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD DISLODGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION, SJ4 CONNECTOR | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120Q/58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Required Intervention |