DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 1895891
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-04507
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- August 30, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE LEAD WAS RECEIVED CUT IN TWO PIECES. SINCE THE LEAD WAS CUT IN THE FIELD, THE LEAD IMPEDANCE AND HIGH VOLTAGE INTEGRITY COULD NOT BE MEASURED. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS. LOW SENSING WAS ALSO OBSERVED. X-RAY REVEALED LEAD DISLODGMENT. LEAD WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |