FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 1895740 · Received November 10, 2010

Report

Report Number
2017865-2010-04623
Event Type
Injury
Date Received
November 10, 2010
Date of Event
August 24, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE CONNECTOR PIN WAS NORMAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH THRESHOLD AND LOW IMPEDANCE WERE OBSERVED. DURING LEAD REPOSITIONING, THE CONNECTOR PIN FELT AS IF IT WAS LOOSENING WITHIN THE CONNECTOR. THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention