FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 1895707 · Received November 10, 2010

Report

Report Number
2017865-2010-04528
Event Type
Injury
Date Received
November 10, 2010
Date of Event
August 19, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NV
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT LOW SENSING, HIGH IMPEDANCE AND HIGH THRESHOLD WERE OBSERVED. ECHOCARDIOGRAM REVEALED LEAD PERFORATION. THE LEAD WAS REPOSITIONED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention