FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 1895686 · Received November 10, 2010

Report

Report Number
2017865-2010-04516
Event Type
Injury
Date Received
November 10, 2010
Date of Event
August 30, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE WAS OBSERVED ON THE VENTRICULAR AND ATRIAL LEAD DUE TO INSULATION DAMAGE. HIGH CAPTURE THRESHOLDS WAS OBSERVED. THE NOISE WAS REPRODUCIBLE ON THE VENTRICULAR LEAD BUT NOT REPRODUCIBLE ON TH E ATRIAL LEAD. THE RV LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/52 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention